Guyana In-situ Raman battery testing cell E704A

1. Breathable, enabling testing under gas‑circulating conditions. 2. Negative electrode separator protection, reducing negative electrode degradation and effectively extending the experimental duration.

Category:

GuyanaIn-situ Raman electrochemical cell

Product Description

In-situ Raman Battery Test Cell E704A — Two-Electrode Configuration, Gas Inlet, and Anode Protection
Material Main material: 316 stainless steel
Sealing component materials: polytetrafluoroethylene, fluoroelastomer
Window material: Quartz glass
Main Parameters

Overall external dimensions: Ø54×19
Window size: Ø5 (Ø8 and Ø10 available)
Sample distance from the window: 2 mm
Window thickness: 0.5 mm
Suitable for mounting sample sizes: Ø12–Ø16 (Ø5 window), Ø15–Ø16 (Ø8 window)

Inner diameter: Ø16

Uses and Features 1. It is breathable and can be tested under gas-circulation conditions.
2. Negative electrode separator protection reduces negative electrode degradation and effectively extends the experimental duration.